Commands Listed in Alphabetical Order
BUS
Sets or returns the parameters for each bus. These parameters affect either the
Serial Trigger Setup or the Bus Display.
Conditions
This command requires a DPO2AUTO, DPO2EMBD, or D PO2COMP application
module.
Group
Bus
Syntax
BUS
BUS:B<x>:CAN:BITRate
Sets or returns the bit rate for CAN bus <x>, where x is the bus number.
Conditions
This command requires a DPO2AUTO applic ation module.
Group
Bus
Syntax
BUS:B<x>:CAN:BITRate <NR1>
BUS:B<x>:CAN:BITRate?
Arguments
<NR1> sets the bit rate to the closest bit rate supported by the instrument.
Returns
The query always returns the numerical bit rate value.
BUS:B<x>:CAN:PRObe
Sets or returns the probing method to probe CAN bus <x>, where x is the bus
number.
Conditions
This command requires a DPO2AUTO applic ation module.
Group
Bus
MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual 2-73